Barrel Switch Troubleshooting - Steam Series
Barrel Switch Troubleshooting
Objective:
Test magnetic barrel switches to diagnose steam solenoid issues.
Time:
15-20 minutes
Tools:
#2 Philips Head Screw Driver
Multimeter
Adjusting Sensitivity
Steam barrel switches are threaded into the housing and are adjustable by screwing in or out. If it's not adjusted properly, it could affect the response of the actuator.
If the steam solenoid isn't engaging when actuating, try unthreading it counter-clockwise so the switch moves further away from the magnet.
If the switch doesn't respond despite adjustment, you'll need to test the functionality.
Testing Functionality
Step 1:
Step 2:
Remove the bottom front panel, being sure to disconnect the LED group lights.
Step 3:
To identify the correct barrel is being used, test the barrel's continuity with a multimeter. Steam barrel switches are N/C (normally closed). It should only be active and have continuity when a magnet is not present. Conversely, there is no continuity when a magnet is present.
Please note, to test this properly you need to remove the barrel from the assembly.
The continuity should reliably change based on the presence of the magnet.
Related Articles
Group Magnet Switch Troubleshooting - Steam Series
Group Mag Switch Troubleshooting Steam LP & Steam X Objective: Inside the group head is a tongue with a magnet installed on it. This magnet comes in contact with the magnetic reed switches which send signals to the control board for brewing. ...
Group Magnet Switch Troubleshooting - Espresso Series
Group Mag Switch Troubleshooting - Espresso V3 & Single Group SLAYER recommends all service be performed by an experienced SLAYER Espresso trained technician. For a technician reference near you, please contact SLAYER Support at (206)-284-7171 (#1 in ...
Pre-Infusion Solenoid Troubleshooting - Steam LP
Pre-Infusion Solenoid Troubleshooting - Steam LP SLAYER recommends all service be performed by an experienced SLAYER Espresso trained technician. For a technician reference near you, please contact SLAYER Support at (206)-284-7171 (#1 in Directory) ...
Reservoir Magnet Switch Bracket - Steam Single
Field technicians guide for E015 fault. SLAYER recommends all service be performed by an experienced SLAYER Espresso trained technician. For a technician reference near you, please contact SLAYER Support at (206)-284-7171 (#1 in Directory) or email ...
Error Code Identification - Steam Single
Steam Single Error Code Identification 015 - Reservoir Low Water Level Machine software detected a condition of low water level in the reservoir. 016 - Brew Actuator Position Error Software detected an electronics failure of actuator switch contact. ...